We have all become very good at optimizing the ocular surface and using the latest biometers. Sometimes it is shocking when you walk in and see a perfectly quiet eye with a clear cornea, but the patient is unhappy seeing 20/30, and you just found out your intraocular lens (IOL) power did not hit your target.
It is frustrating because the surgery went perfectly. All the measurements were done perfectly. And now you have to explain to your patient who spent thousands of dollars on their lens implant why they’re not seeing as well as they hoped.
That is the uncomfortable reality of premium cataract surgery.
You would think that refractive accuracy should have dramatically improved after buying the latest swept-source optical coherence tomography biometry, modern corneal tomography, total keratometry, and using the new generation of IOL formulas. The technology available to us today would have seemed extraordinary when I started operating.
But despite all of this planning with the best measurements possible, a perfectly executed surgery, and a quiet eye, the patient misses their refractive target and is unhappy.
We all know that every eye has its own biology, and some of that biology remains unpredictable. But now we have to explain that to the patient in front of you. I remember telling a patient once, “I replaced this lens because all the formulas said this was the correct lens, but it turns out this other lens would’ve been better.” And I was trying to explain this, but the only message the patient received was, “So you put the wrong lens in my eye.”
For a monofocal patient with reasonable expectations, half a diopter may simply be a conversation. For a premium-lens patient who paid several thousand dollars specifically to reduce dependence on glasses, the same half diopter can become a much bigger issue.
So after years of trying to minimize this problem, communicate more efficiently, and have better conversations with my patients, I wrote this article to give you some pearls.
The first goal is obvious: Reduce the chance of refractive error as much as possible.
The second receives far less attention, but I think it is just as important: Design the refractive plan so that when a miss occurs, it does the least possible damage to the patient's overall vision.
That second goal has changed the way I approach premium cataract surgery.
Good Formulas Cannot Rescue Bad Data
Before arguing about which IOL formula is best, I want to know that the numbers going into the formula are worth trusting, and this starts with the ocular surface.
If a premium-lens candidate has significant dry eye or epithelial irregularity, I treat the surface first to make sure the cornea will be able to give good readings. Depending on the patient, that may involve preservative-free lubrication, brief steroid treatment, management of lid disease, and enough time for the cornea to stabilize before repeating measurements.
Hovanesian and colleagues demonstrated the principle nicely. In cataract patients with dry eye, biometry repeated after 28 days of lifitegrast 5% twice daily produced more accurate postoperative refractive predictions than measurements obtained before treatment.1
So even if we have the most advanced biometer in the world, the readings are going to be inaccurate if we don’t have a good ocular surface. In that same light, it’s important to make sure the patient is not using contact lenses for one to two weeks prior to biometry; they should be removed longer for hard contacts. It is important to determine whether the cornea has actually stopped changing after discontinuing the contact lenses. The lenses should remain out until the measurements are repeatable.
Today's biometers are remarkable. The IOLMaster 700 (Zeiss), Eyestar 900 (Haag-Streit), ARGOS (Alcon), and other modern platforms provide highly reproducible measurements and can obtain very precise and detailed data, allowing us greater ability to predict lens power, astigmatism, power, and astigmatism direction.
But owning sophisticated equipment is not the same as using it critically. For every cataract case, I compare axial length, keratometry, astigmatic magnitude and axis, and the overall corneal pattern with topography and autokeratometry.
When the biometer and topographer disagree, and I cannot explain why, I do not average the two. Rather, I repeat the measurements.
This can sometimes expose the real issue. The patient might have an unstable tear film, residual contact-lens effect, poor fixation, epithelial basement membrane dystrophy, or another irregularity.
It is not important to just obtain the maximum number of readings, but rather to arrive at measurements that tell one coherent story about the eye.
Modern IOL Formulas
Once I am comfortable with the biometry, I want to see what several contemporary formulas predict.
The European Society of Cataract and Refractive Surgeons calculator (iolcalculator.escrs.org) is one of the tools I use frequently because it places Barrett Universal II, Cooke K6, EVO 2.0, Hill-RBF, Hoffer QST, Kane, and PEARL-DGS side by side.3
The Ladas Super Formula (iolcalc.com) is another useful option—and it was using artificial intelligence before it was in fashion.
It is helpful to see if the IOL values agree formula to formula. When several strong formulas cluster tightly around the same result, I am reassured. When the predictions spread widely, I stop and ask why.
The disagreement is often telling me something about the eye rather than something about the software. However, sometimes the formulas are going to disagree by as much as 0.5 diopters. Given this possible range of results, I then formulate a plan to see which range would be the most ideal for the given patient.
For post-LASIK, post-photorefractive keratectomy, and post-radial keratotomy eyes, I also use tools designed specifically for post-refractive surgery calculations, including the calculator from the American Society of Cataract and Refractive Surgery. These eyes are simply less predictable, and I discuss that reality with patients before surgery. Barrett True-K has performed comparably with multiple-formula approaches in a published series.4
It is always a good idea to document that you reviewed with your patient increase risk in IOL power error after corneal surgery.
For toric IOLs, I use the calculator recommended for the specific lens platform. Most lens manufacturers' toric calculators require anterior corneal measurements rather than total keratometry. Many of the modern biomes that calculate posterior corneal astigmatism will provide an additional recommendation on toric power and alignment as well. Given all this, it is surprising how inaccurate our outcomes can be.
Before quoting refractive accuracy to patients, I think we should be honest with ourselves about the refractive accuracy, using all the best methods to prevent an IOL power error.
Calculation performance of IOLs is usually described by the percentage of eyes landing within a given range of the predicted postoperative spherical equivalent. (These numbers are not universal, as outcomes vary with patient population, lens, biometer, formula, constants, surgeon, and study design.)
In one representative series of 324 eyes measured with the IOLMaster 700, the Barrett Universal II formula yielded prediction errors within ±0.25 D in approximately 42.7% of eyes, within ±0.50 D in 79.4%, within ±0.75 D in 92.4%, and within ±1.00 D in 97.2%.5 In more recent studies using current-generation formulas, top-performing methods have placed approximately 86% to 87.5% of eyes within ±0.50 D.6
But let’s turn that statistic around.
Despite all the careful planning that goes into our IOL power selection, slightly over 10% of our patients will still finish more than half a diopter from target despite careful measurements and modern formulas.
These highly controlled trial data make an important point.
In the prospective randomized clinical trial of Alcon’s AcrySof IQ Vivity IOL, investigators targeted emmetropia. At six months, 91.6% of first eyes and 86.8% of second eyes were within ±0.50 D of emmetropia. Among 213 Vivity eyes, 23 (10.8%) finished more than 0.50 D from emmetropia.7 This was a controlled clinical trial with standardized protocols and carefully selected investigators, yet roughly 1 in 10 eyes still ended up more than half a diopter from plano. To me, that is an important reality check.
Uncertainty in the Labeled IOL
Another source of variability receives surprisingly little discussion: the lens itself.
Under the ISO 11979-2:2024 standard recognized by the US Food and Drug Administration, allowable deviation from labeled spherical power increases with IOL power:
-
±0.30 D at 15 D or below
-
±0.40 D above 15 D through 25 D
-
±0.50 D above 25 D through 30 D
-
±1.00 D above 30 D8,9
That deserves some perspective.
We refine biometry, model posterior corneal astigmatism, optimize constants, and debate refractive targets in quarter-diopter increments. Meanwhile, the power printed on the IOL package is a nominal labeled value within an accepted manufacturing tolerance.
Langenbucher and colleagues examined the potential effect using Monte Carlo analysis based on 16,669 biometric datasets. They estimated spectacle-plane refractive uncertainty attributable to International Organization for Standardization (ISO) labeling tolerance at roughly 0.18 to 0.19 D, with greater influence in eyes requiring higher-power IOLs.10
That does not mean modern IOL manufacturing is imprecise. Quite the opposite: The ISO limits represent maximum allowable deviations, not the amount of deviation we should expect from every lens.
A 2025 ISO-compliant optical bench study reported average absolute deviations of 0.137 D for the toric components tested and 0.057 D for multifocal components, with the tested lenses meeting ISO standards overall.11
Another 2025 clinical series, in which the manufacturer supplied the exact measured power of each implanted lens, found an average difference between exact and labeled power of 0.12 D for lenses from 15 to 25 D and 0.19 D for lenses from 25 to 30 D. Interestingly, substituting exact lens power for labeled power did not meaningfully improve postoperative prediction in that study.12
This is not an argument that IOL manufacturing is sloppy. It is simply another reminder that the number printed on the box is not perfect and is more variable the higher the power of the lens.
So if we know that 1 out of 10 of our premium lens patients is likely to have a significant IOL power error, what are we going to do about it?
Widen the Landing Zone
This is where my own practice has evolved.
In appropriately selected patients, I increasingly use an extended depth of focus (EDOF) mini-monovision strategy, generally beginning with the nondominant eye.
With lenses such as TECNIS PureSee (Johnson & Johnson Vision) and Vivity (Alcon), I commonly target modest myopia in that eye, often around -0.5 to -1.0 D.13-15 I have found the nondominant eye allows more room for error in the IOL power if this eye is used for mild amounts of myopia. I individualize the target for each patient. Usually this falls somewhere between approximately -0.5 and -1.0 D depending on occupation, visual priorities, reading habits, previous monovision experience, binocular status, and tolerance for anisometropia.15,16 Someone who prioritizes crisp distance and is comfortable using readers may stay closer to the distance end of that range. Someone who places greater value on functional near vision may reasonably accept somewhat more myopia.
Let the First Eye Teach You About the Second
The first operated eye may be the most valuable patient-specific data point we get.
Once the eye is quiet and the refraction is repeatable, I evaluate the spherical equivalent, residual cylinder, and acuity at distance, intermediate, and near.
It is important to stop staring at the numbers and talk to the patient and ask the patient:
-
How is your distance vision?
-
How is your computer vision?
-
How is your reading?
-
What can you do without glasses now that you could not do before?
-
What do you wish were better?
-
Are you noticing halos or glare at night?
Those answers often tell me more than the manifest refraction alone.
“I can read everything, but I wish the television were sharper,” is a very different problem from “Distance is fantastic, but I thought my phone would be easier.” These two patients may have similar refractions and completely different priorities for the second eye.
The first eye can also reveal important information about the patient’s unique biology, which can help guide a more accurate IOL power prediction for the other eye.
There is published support for using first-eye prediction error to refine second-eye planning when the two eyes are reasonably similar. Importantly, the full first-eye error should not simply be transferred blindly. Formula-specific partial adjustments have produced better second-eye predictions than assuming 100% of the error will repeat.17 I treat the first-eye result as another patient-specific input and recalculate.
For the dominant eye, I generally want excellent distance vision. Depending on the result of the first eye and what the patient tells me, that may mean plano or a small amount of myopia.
Do Not Forget the Cylinder
The same philosophy applies to astigmatism.
There is uncertainty in astigmatic correction whether we use limbal relaxing incisionsor toric IOLs. After the first eye, I look at the postoperative refraction and, when appropriate, repeat corneal measurements to understand what the astigmatic treatment actually accomplished in that individual patient: Not what the nomogram predicted, but what actually happened.
If the first eye was overcorrected, that information can influence the second-eye plan. If it was undercorrected, that matters, too. In effect, the first eye gives you a personalized data point that is more relevant to the fellow eye than any population average.
Sequential surgery also creates another opportunity. Residual astigmatism in the first eye can sometimes be addressed when the patient returns for the second eye or later in the office, depending on the clinical situation and the chosen method of enhancement.
Two staged eyes give you information and options that simultaneous bilateral surgery does not.
Mini-Monovision Is Not Traditional Monovision
When some surgeons hear “monovision,” they picture a large interocular difference, often around 1.5 D or more. That can provide strong uncorrected near vision in the myopic eye, but it may also compromise distance vision in that eye, stereoacuity, and comfortable binocular function. Intermediate vision can become an issue as well.
Mini-monovision uses a smaller offset: In a prospective comparison of 0.75 D versus 1.75 D of anisometropia, stereoacuity was significantly better with the smaller difference.18
Extended depth of focus optics add another layer because the mildly myopic eye maintains a broader functional range than a single-focus optic at the same spherical equivalent.15,16
That is the concept I am trying to create: Not one distance eye and one near eye. Two useful eyes with overlapping ranges, with one shifted modestly toward near.
The overlap is the point; it is what makes the system forgiving.
What If the First Eye Misses Anyway?
Suppose I intended the nondominant eye to finish around -0.60 D and it lands at -1.00 D. Before calling that a failure, I ask how the patient is functioning. If near and intermediate vision are excellent and the patient is thrilled, I now have useful information rather than an automatic problem. I can use the second eye to protect distance.
Conversely, if the first eye lands closer to plano than expected and the patient loves the distance but wants more reading ability, the second eye gives me an opportunity to rebalance the binocular result toward near.
The postoperative number still matters; it just does not get to define success by itself. That is the practical payoff.
In premium cataract surgery, the endpoint is not a spherical equivalent sitting by itself in the chart. The endpoint is binocular function that matches what the patient told you they wanted.
Dominance Is a Data Point, Not a Law
I still test ocular dominance, and I still generally prefer the nondominant eye for the myopic target. But I think of dominance as useful information rather than an unbreakable rule.
Crossed monovision has produced satisfactory results in selected patients, with published series reporting visual function, spectacle independence, and satisfaction comparable with conventional monovision.19 Once the first eye has been done, what the patient actually experiences may be more useful than a dominance test performed weeks earlier.
That is another advantage of sequential surgery. It gives you information no formula can generate: how this particular patient's visual system responds to this particular optical plan.
Have This Conversation Before Surgery
None of this works well if the patient hears about refractive uncertainty for the first time when they are unhappy after surgery.
I discuss it during the consultation, explaining that we use sophisticated measurements, repeat anything questionable, optimize the ocular surface, compare modern formulas, and do everything reasonable to improve refractive accuracy. I also explain that we cannot perfectly predict how an individual eye will heal or exactly where an IOL will ultimately sit.
Occasionally, an enhancement becomes part of the overall refractive plan. That conversation sounds very different before surgery than it does afterward.
Before surgery, it is expectation setting and informed consent. After an unexpected result, the same explanation can sound like an excuse.
No surgeon wants to hear, “Doctor, you put the wrong lens in my eye.” The best way to reduce that reaction is not to promise perfection; it is to explain the uncertainty honestly and have a plan for dealing with it.
The Real Goal Is a Bigger Landing Zone
Biometers will continue to improve. Formulas will improve. Artificial intelligence models, posterior corneal measurements, and effective-lens-position prediction will improve. Our refractive accuracy should improve with them.
I am not arguing against any of that.
I am arguing against building a premium cataract strategy around the assumption that the calculation will be right every time.
The data tell us it will not be, so I try to do both things: Push the calculation as far as it can reasonably go, then build a binocular optical plan with enough flexibility to tolerate the fact that biology occasionally refuses to cooperate.
Optimize the surface before measuring. Repeat measurements that do not make sense. Compare contemporary formulas and treat disagreement as useful information rather than an inconvenience. Use the first eye to learn something about the second. And, in the right patient, consider an EDOF mini-monovision strategy that creates overlapping binocular ranges instead of two narrow focal points sitting beside each other.
Because ultimately, the goal is not simply to hit a number. The goal is to give the patient a visual system that still works well when the number is not exactly the one you predicted.
The IOL power error is going to happen. Our job is to make it happen less often, and to make it matter less when it does.
References
-
Hovanesian JA, Epitropoulos A, Donnenfeld ED, Holladay JT. The effect of lifitegrast on refractive accuracy and symptoms in dry eye patients undergoing cataract surgery. Clin Ophthalmol. 2020;14:2709-2716. doi:10.2147/OPTH.S264520
-
Kane JX, Chang DF. Intraocular lens power formulas, biometry, and intraoperative aberrometry: a review. Ophthalmology. 2021;128(11):e94-e114. doi:10.1016/j.ophtha.2020.08.010
-
Skrzypecki J, Koch DD, Wang L. Performance of formulas included in the ESCRS intraocular lens power calculator. J Cataract Refract Surg. 2024;50(12):1224-1229. doi:10.1097/j.jcrs.0000000000001531
-
Ferguson TJ, Downes RA, Randleman JB. IOL power calculations after LASIK or PRK: Barrett True-K biometer-only calculation strategy yields equivalent outcomes as a multiple formula approach. J Cataract Refract Surg. 2022;48(7):784-789. doi:10.1097/j.jcrs.0000000000000883
-
Kim SY, Lee SH, Kim NR, Chin HS, Jung JW. Accuracy of intraocular lens power calculation formulas using a swept-source optical biometer. PLoS One. 2020;15(1):e0227638. doi:10.1371/journal.pone.0227638
-
Voytsekhivskyy OV, Hoffer KJ, Cooke DL, Savini G. IOL power calculation project: accuracy of 36 formulas. Am J Ophthalmol. 2025;277:45-56. doi:10.1016/j.ajo.2025.05.004
-
Alcon Laboratories, Inc. Clareon Vivity Toric extended vision UV absorbing posterior chamber intraocular lens: product information. Alcon Laboratories, Inc; 2022. 900371484-0822.
-
International Organization for Standardization. Ophthalmic implants—Intraocular lenses—Part 2: Optical properties and test methods. 3rd ed.ISO 11979-2:2024. International Organization for Standardization; 2024. https://www.iso.org/standard/86607.html
-
US Food and Drug Administration. Recognized consensus standards: medical devices. ISO 11979-2 third edition 2024-10, Recognition No. 10-137. Accessed August 14, 2026. https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/detail.cfm?standard__identification_no=46072
-
Langenbucher A, Szentmary N, Cayless A, Bolz M, Hoffmann P, Wendelstein J. Prediction of spectacle refraction uncertainties with discrete IOL power steps and manufacturing tolerances according to ISO using a Monte Carlo model. Br J Ophthalmol. 2024;108(6):793-800. doi:10.1136/bjo-2023-323921
-
Han H, Koh JW. Evaluation of optical power integrity in toric and multifocal intraocular lenses. BMC Ophthalmol. 2025;25:565. doi:10.1186/s12886-025-04380-6
-
Fernandez J, Ribeiro F, Burguera N, Rodriguez-Calvo-de-Mora M, Rodriguez-Vallejo M. Influence of manufacturing tolerance and formula thickness type in the prediction error of multifocal intraocular lens power calculation. J Refract Surg. 2025;41(9):e943-e949. doi:10.3928/1081597X-20250707-03
-
Corbett D, Black D, Roberts TV, et al. Quality of vision clinical outcomes for a new fully-refractive extended depth of focus intraocular lens. Eye (Lond). 2024;38(suppl 1):9-14. doi:10.1038/s41433-024-03039-8
-
Black DA, Bala C, Alarcon A, Vilupuru S. Tolerance to refractive error with a new extended depth of focus intraocular lens. Eye (Lond). 2024;38(suppl 1):15-20. doi:10.1038/s41433-024-03040-1
-
Vasavada V, Vasavada SA, Shastri L, Nath V, Vasavada AR, Srivastava S. Visual outcomes comparing emmetropia vs mini-monovision after bilateral implantation of a nondiffractive extended vision intraocular lens: randomized trial. J Cataract Refract Surg. 2024;50(8):799-804. DOI: 10.1097/j.jcrs.0000000000001458
-
Fernandes MC, Nunomura CY, Messias A. Visual performance and photic disturbances with diffractive and nondiffractive EDOF intraocular lenses using mini-monovision: randomized trial. J Cataract Refract Surg. 2024;50(2):153-159. doi:10.1097/j.jcrs.0000000000001330
-
Turnbull AMJ, Barrett GD. Using the first-eye prediction error in cataract surgery to refine the refractive outcome of the second eye. J Cataract Refract Surg. 2019;45(9):1239-1245. doi:10.1016/j.jcrs.2019.04.008
-
Hayashi K, Ogawa S, Manabe S, Yoshimura K. Binocular visual function of modified pseudophakic monovision. Am J Ophthalmol. 2015;159(2):232-240. doi:10.1016/j.ajo.2014.10.023
-
Zhang F, Sugar A, Arbisser L, Jacobsen G, Artico J. Crossed versus conventional pseudophakic monovision: patient satisfaction, visual function, and spectacle independence. J Cataract Refract Surg. 2015;41(9):1845-1854. doi:10.1016/j.jcrs.2015.10.013